Z-0171-2025 Class II Ongoing
FDA device recall Z-0171-2025 was initiated by MEDLINE INDUSTRIES, LP - Northfield on Sep 4, 2024 and is designated Class II. Reason for recall: It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is… The recall status is ongoing. Affected quantity: 194 units.
Recall Details
- Product Type
- Devices
- Report Date
- Oct 30, 2024
- Initiation Date
- Sep 4, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 194 units
Product Description
Medline brand medical procedure convenience kits, labeled as: a) UNIVERSAL DRAPE PACK, REF DYNJ44114; b) DRAPE PACK, REF DYNJ82663B
Reason for Recall
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Canada, Panama.
Code Information
a) REF DYNJ44114, UDI/DI 40888277202055 (each), 10888277202054 (case), Lot Numbers: 24ABB398; b) REF DYNJ82663B, UDI/DI 40195327544325 (each), 10195327544324 (case), Lot Numbers: 23LBV498