Z-0185-2025 Class II Ongoing

Recalled by Smiths Medical ASD, Inc. — Minneapolis, MN

FDA device recall Z-0185-2025 was initiated by Smiths Medical ASD, Inc. on Sep 24, 2024 and is designated Class II. Reason for recall: Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar… The recall status is ongoing. Affected quantity: 400 units.

Recall Details

Product Type
Devices
Report Date
Oct 30, 2024
Initiation Date
Sep 24, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
400 units

Product Description

medex TRANSTAR MONITORING KIT 1/EA, Product Code REF MX20103R1

Reason for Recall

Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

Distribution Pattern

Worldwide distribution.

Code Information

UDI/DI 50351688551694 (case), 10351688551696 (pouch), Lot Numbers: 4270117