Z-0194-2023 Class II Ongoing

Recalled by Medtronic MiniMed — Northridge, CA

FDA device recall Z-0194-2023 was initiated by Medtronic MiniMed on Sep 20, 2022 and is designated Class II. Reason for recall: Medtronic identified a cybersecurity vulnerability in the MiniMed 600 series Insulin Pump Systems associated with the communication protocol that could allow unauthorized access to the pump system… The recall status is ongoing. Affected quantity: 316326.

Recall Details

Product Type
Devices
Report Date
Nov 16, 2022
Initiation Date
Sep 20, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
316326

Product Description

Insulin Pump/Model: MiniMed 620G/ MMT-1750 MiniMed 640G/ MMT-1711, MMT-1712, MMT-1751, MMT-1752

Reason for Recall

Medtronic identified a cybersecurity vulnerability in the MiniMed 600 series Insulin Pump Systems associated with the communication protocol that could allow unauthorized access to the pump system. This unauthorized access could be used to deliver too much or too little insulin through delivery of an unintended insulin bolus or because insulin delivery is slowed or stopped, which could lead to hypoglycemia or hyperglycemia. The Remote Bolus feature on insulin pumps should be turned off to prevent the unauthorized access.

Distribution Pattern

US: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY. OUS: Andorra, Australia, Austria, Barbados, Belgium, Bermuda, Bolivia, Botswana, Canada, Canary Islands, Cayman Islands, Chile, Czech Republic, Denmark, Dominican Rep., El Salvador, Faroe Islands, Finland, France, Germany, Gibraltar, Greece, Guatemala, Hungary, Ireland, Israel, Italy, Kenya, Liechtenstein, Luxembourg, Mauritius, Mexico, Namibia, Netherlands, New Zealand, Norway, Paraguay, Peru, Poland, Portugal, Romania, Slovakia, South Africa, Spain, Sweden, Switzerland, Trinidad and Tobago, Turkey, Uganda, United Kingdom, Venezuela

Code Information

Model/UDI-GTIN (All Serial Numbers): MMT-1750/00763000375461, 00643169559745; MMT-1711/643169554931, 763000367039, 00643169890039, 00643169554917, 00643169629813, 00643169554924, 00643169554931, 00643169742062, 00763000013066, 00643169554948, 00643169554955, 00643169621954; MMT-1712/643169662612, 643169577701, 00763000205409, 00643169890046, 00643169577664, 00643169629820, 00643169577671, 00643169577688, 00763000155346, 00643169577695, 00643169577701, 00643169621961; MMT-1751/643169672239, 00643169574410, 00643169574427, 00643169574434, 00763000253288, 00643169574441, 00643169574458, 00643169643512, 00643169521155, 00643169521704, 00763000015596, 00643169520882, 00643169521421, 00643169521971; MMT-1752/763000192181, 00643169596368, 00643169596382, 00643169596405, 00763000318291, 00763000318307, 00643169596443, 00643169522305, 00643169521292, 00643169522381, 00643169521841, 00643169522268, 00643169521025, 00643169522343, 00643169521575, 00643169522114

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0194-2023
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 24, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.