Z-0194-2025 Class II Ongoing

Recalled by Smiths Medical ASD, Inc. — Minneapolis, MN

FDA device recall Z-0194-2025 was initiated by Smiths Medical ASD, Inc. on Sep 24, 2024 and is designated Class II. Reason for recall: Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar… The recall status is ongoing. Affected quantity: 1000 units.

Recall Details

Product Type
Devices
Report Date
Oct 30, 2024
Initiation Date
Sep 24, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1000 units

Product Description

medex TranStar DPT w/Stopcock & 30cc Flush 10/EA, Product Code REF MX950SCNT

Reason for Recall

Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

Distribution Pattern

Worldwide distribution.

Code Information

UDI/DI 50351688503938 (case), 10351688503930 (pouch), Lot Numbers: 4293126, 4309186, 4316649, 4323123, 4328467,4346049

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0194-2025
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.