Z-0207-2013 Class II Terminated

Recalled by OrthoPediatrics Corp — Warsaw, IN

FDA device recall Z-0207-2013 was initiated by OrthoPediatrics Corp on Sep 13, 2012 and is designated Class II. Reason for recall: A complaint received from sales representative that a replenishment instrument was too short. Upon investigation, it has been determined that the device was not manufactured to design specifications… The recall status is terminated (terminated Sep 18, 2013). Affected quantity: 78.

Recall Details

Product Type
Devices
Report Date
Nov 14, 2012
Initiation Date
Sep 13, 2012
Termination Date
Sep 18, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
78

Product Description

Trochar Scalpel Instrument (Surgical Punch) OrthoPediatrics intramedullary rods (nails) are generally rod-shaped devices, with screw holes at either end for fixation to bone. This device is intended to be inserted into the medullary canal of the femur for fixation of fractures by aligning and stabilizing the bone fragments. Additional stabilization may be realized by installing transverse screws through holes in the rod. These devices are made of medical grade stainless steel. The OrthoPediatrics PediNail" system is used for pediatric and small stature adult patients as indicated to stabilize fractures of the femoral shaft; subtrochanteric fractures; ipsilateral neck/shaft fractures; prophylactic nailing of impending pathologic fractures; nonunions and malunions; fixation of femurs that have been surgically prepared (osteotomy) for correction of deformity.

Reason for Recall

A complaint received from sales representative that a replenishment instrument was too short. Upon investigation, it has been determined that the device was not manufactured to design specifications and will not function as intended.

Distribution Pattern

Nationwide Distribution including AL, IL, NY, FL, TX, MI, KY, NC, CO, and GA.

Code Information

Part number: 01-1500-9014 and Lot #'s 1466707 and 1466710

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0207-2013
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.