Z-0209-2024 Class II Ongoing

Recalled by Abbott — Plymouth, MN

FDA device recall Z-0209-2024 was initiated by Abbott on Sep 28, 2023 and is designated Class II. Reason for recall: Certain NT2000iX RF generators were serviced using a tool that was subsequently found to be out of tolerance and therefore may not have provided accurate results for potential capacitance failure. The recall status is ongoing. Affected quantity: 34 devices.

Recall Details

Product Type
Devices
Report Date
Nov 8, 2023
Initiation Date
Sep 28, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
34 devices

Product Description

Abbott NT2000iX Radiofrequency (RF) generator, REF RFG-NT-2000, lesioning generator

Reason for Recall

Certain NT2000iX RF generators were serviced using a tool that was subsequently found to be out of tolerance and therefore may not have provided accurate results for potential capacitance failure.

Distribution Pattern

US, Taiwan, and Colombia

Code Information

UDI/DI 05415067022417, serial numbers: 16930509, 16983330, 17583407, 17634837, 17783928, 17783930, 17905961, 17928413, 17954521, 17991999, 18047384, 18493711, 18522937, 18675068, 18690920, 18709734, 18742997, 2475-12, 2504-12, 2547-13, 2885-14, 3003-14, 3278-15, 3450-15, 3550-15, 17942357, 18003310, 18102335, 18132017, 18132092, 18657521, 18666647, 19182119, 19197174

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0209-2024
openFDA file date
Nov 3, 2026
Loaded by RegDataLab
Nov 4, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0209-2024. Report a data issue: hello@regdatalab.com.