Z-0209-2024 Class II Ongoing

Recalled by Abbott — Plymouth, MN

Recall Details

Product Type
Devices
Report Date
November 8, 2023
Initiation Date
September 28, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
34 devices

Product Description

Abbott NT2000iX Radiofrequency (RF) generator, REF RFG-NT-2000, lesioning generator

Reason for Recall

Certain NT2000iX RF generators were serviced using a tool that was subsequently found to be out of tolerance and therefore may not have provided accurate results for potential capacitance failure.

Distribution Pattern

US, Taiwan, and Colombia

Code Information

UDI/DI 05415067022417, serial numbers: 16930509, 16983330, 17583407, 17634837, 17783928, 17783930, 17905961, 17928413, 17954521, 17991999, 18047384, 18493711, 18522937, 18675068, 18690920, 18709734, 18742997, 2475-12, 2504-12, 2547-13, 2885-14, 3003-14, 3278-15, 3450-15, 3550-15, 17942357, 18003310, 18102335, 18132017, 18132092, 18657521, 18666647, 19182119, 19197174