Z-1480-2024 Class II Ongoing

Recalled by Abbott GmbH — Wiesbaden

Recall Details

Product Type
Devices
Report Date
April 10, 2024
Initiation Date
March 4, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5110 units

Product Description

Alinity i Anti-HCV Reagent Kit, List Number 08P0521

Reason for Recall

Customer complaints were received regarding falsely elevated results for some patient samples. Internal studies with customer return samples confirmed potential interactions that may lead to falsely elevated results when processed as below on the same instrument: " ARCHITECT Syphilis TP (LN 8D06) and/or ARCHITECT AdviseDx SARS-CoV-2-IgG II (LN 6S60) precedes ARCHITECT Anti-HCV testing (LN 1L79). " Alinity i Syphilis TP (LN 07P60) precedes Alinity i Anti-HCV testing (LN 08P05).

Distribution Pattern

US Nationwide distribution.

Code Information

UDI/DI 00380740130657, Lot Numbers: 52020BE00, 54149BE00, 56384BE00, 58172BE00, 59203BE00