Z-0212-2016 Class I Terminated

Recalled by Boston Scientific Corporation — Maple Grove, MN

FDA device recall Z-0212-2016 was initiated by Boston Scientific Corporation on Oct 9, 2015 and is designated Class I. Reason for recall: Boston Scientific is recalling its recently released RotaWire Elite Guidewire and wireClip Torquer Guidewire Manipulation device that is used in conjunction with the Rotablator Rotational Atherectomy… The recall status is terminated (terminated Oct 13, 2017). Affected quantity: 955 single units (191 5-packs).

Recall Details

Product Type
Devices
Report Date
Nov 18, 2015
Initiation Date
Oct 9, 2015
Termination Date
Oct 13, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
955 single units (191 5-packs)

Product Description

RotaWire Elite and wireClip Torquer, Guidewire and Guidewire Manipulation Device, Extra Support. REF H802233301 (5-pack outer package), REF H802233300 (single unit). The Rotablator rotational atherectomy system consists of a diamond tipped burr capable of high speed rotation designed for plaque modification and debulking in vascular interventions. This system is designed to travel over a specialized 0.009 guidewire to guide the path of burr travel to deliver therapy where intended. The guidewires are intended for use with the Rotablator Rotational Atherectomy System.

Reason for Recall

Boston Scientific is recalling its recently released RotaWire Elite Guidewire and wireClip Torquer Guidewire Manipulation device that is used in conjunction with the Rotablator Rotational Atherectomy System because of complaints of wire fracture.

Distribution Pattern

Worldwide Distribution: US Distribution in the states of AZ, CA, DE, FL, GA, IL, LA, MD, MI, MN, NE, NY, PA, SC, TX, WA, and WI, and the countries of Belgium, Germany, Great Britain, Italy, Japan, and South Africa.

Code Information

Lots: 18141187, 18194620, 18220238, 18261834, 18285502, 18303265, 18390089

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0212-2016
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.