Z-0232-2016 Class II Terminated

Recalled by Boston Scientific Corporation — Saint Paul, MN

FDA device recall Z-0232-2016 was initiated by Boston Scientific Corporation on Sep 17, 2015 and is designated Class II. Reason for recall: Devices shipped to US in dual-channel rather than US approved single-channel RF Telemetry communication mode. While this does not impact therapy delivery of device; this can impact likelihood of… The recall status is terminated (terminated Mar 8, 2016). Affected quantity: 3.

Recall Details

Product Type
Devices
Report Date
Nov 18, 2015
Initiation Date
Sep 17, 2015
Termination Date
Mar 8, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3

Product Description

Boston Scientific SQ-RX Model 1010 subcutaneous pulse generator, (S-ICD). Intended to provide defibrillation therapy.

Reason for Recall

Devices shipped to US in dual-channel rather than US approved single-channel RF Telemetry communication mode. While this does not impact therapy delivery of device; this can impact likelihood of successfully establishing telemetry during first scan of device when using a single-channel programmer. Telemetry can be established with additional scans; once established no further difficulty is expecte

Distribution Pattern

Distributed in the US to MA and PR.

Code Information

A019995 A020132 A020240

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0232-2016
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026