Z-0232-2023 Class II Ongoing

Recalled by Becton Dickinson & Co. — Sparks, MD

FDA device recall Z-0232-2023 was initiated by Becton Dickinson & Co. on Oct 13, 2022 and is designated Class II. Reason for recall: Misformulated minocycline-can cause erroneously increased Minimal Inhibitory Concentration (MIC) values when performing antibiotic susceptibility testing (AST) against minocycline and result in the… The recall status is ongoing. Affected quantity: 13,900 EA.

Recall Details

Product Type
Devices
Report Date
Nov 23, 2022
Initiation Date
Oct 13, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13,900 EA

Product Description

BD Phoenix PMIC/ID-107: The Phoenix PMIC/ID panel is used for the rapid identification and susceptibility of most aerobic and facultative anaerobic gram-positive bacteria of human origin with selected antimicrobial agents when used with the BD Phoenix Automated Microbiology System. Catalog (Ref) Number: 448607

Reason for Recall

Misformulated minocycline-can cause erroneously increased Minimal Inhibitory Concentration (MIC) values when performing antibiotic susceptibility testing (AST) against minocycline and result in the reporting of false resistance for clinical samples if the initial Quality Control (QC) testing of the panel batch passes, delaying treatment of the patient with the most appropriate antibiotic

Distribution Pattern

US Nationwide distribution.

Code Information

UDI: (10) 1355046 (17) 230131 (30) 0025 (01) 30382904486076 Lot Number: 1355046 Exp. Date: 31JAN2023

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0232-2023
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 24, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.