Z-0237-2018 Class II Terminated

Recalled by Beckman Coulter Inc. — Brea, CA

FDA device recall Z-0237-2018 was initiated by Beckman Coulter Inc. on November 3, 2017 and is designated Class II. Reason for recall: The shield for the Single-Tube Station on the DxH instrument may interfere with sample aspiration in the open vial position. The recall status is terminated (terminated June 19, 2018). Affected quantity: 1 unit.

Recall Details

Product Type
Devices
Report Date
December 20, 2017
Initiation Date
November 3, 2017
Termination Date
June 19, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Product Description

UniCel DxH600, Part Number B23858, automated hematology analyzer (Note: the DxH 600 is the same instrument as the DxH 800 except that it is a bench top unit with no instrument cart, computer stand, and no connectivity capabilities)

Reason for Recall

The shield for the Single-Tube Station on the DxH instrument may interfere with sample aspiration in the open vial position.

Distribution Pattern

CA, AR, HI, CA, OH, WA, MD, Canada, Japan

Code Information

Serial Number AZ06682