Z-0244-2016 Class II Terminated

Recalled by Cook Inc. — Bloomington, IN

FDA device recall Z-0244-2016 was initiated by Cook Inc. on Oct 21, 2015 and is designated Class II. Reason for recall: Cook Medical has received six product complaints associated with reports of the 2.6FR curved catheter being mixed with the 4.0FR straight catheter and vice versa. Investigation revealed the products… The recall status is terminated (terminated Nov 28, 2016). Affected quantity: 38.

Recall Details

Product Type
Devices
Report Date
Nov 18, 2015
Initiation Date
Oct 21, 2015
Termination Date
Nov 28, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
38

Product Description

Cook Medical CXI Support Catheter 2.6FR curved catheter and 4.0FR straight catheter, for cardiovascular use.

Reason for Recall

Cook Medical has received six product complaints associated with reports of the 2.6FR curved catheter being mixed with the 4.0FR straight catheter and vice versa. Investigation revealed the products were mixed during processing.

Distribution Pattern

Worldwide Distribution - US Distribution to the states of : CA, FL, ID, IL, MA, NY, OK, TX, UT and VA., and to the countries of Canada, Austria, Germany, Spain, France, Hungary, Sweden and UK.

Code Information

Catalog number: CXI-2.6-18-90-P-NS-ANG Lot: 5505259 Expiration date: December 10, 2017 Catalog number: CXI-4.0-35-90-P-NS-0 Lots: 5505279 and 5505279X Expiration date: December 10, 2017

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0244-2016
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.