Z-0292-2026 Class II Ongoing

Recalled by GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING — Haifa

Recall Details

Product Type
Devices
Report Date
October 29, 2025
Initiation Date
September 12, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
45 units

Product Description

Elscint SPX4; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Reason for Recall

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Distribution Pattern

Worldwide

Code Information

all serial numbers in distribution