Z-0317-2017 Class II Terminated

Recalled by Zimmer Biomet, Inc. — Warsaw, IN

FDA device recall Z-0317-2017 was initiated by Zimmer Biomet, Inc. on Sep 28, 2016 and is designated Class II. Reason for recall: Complaints have been reported that the handpiece is inoperable. The investigation found that the complaints were related to the lack of preventative maintenance contrary to what is recommended in the… The recall status is terminated (terminated Apr 16, 2020). Affected quantity: 703.

Recall Details

Product Type
Devices
Report Date
Nov 2, 2016
Initiation Date
Sep 28, 2016
Termination Date
Apr 16, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
703

Product Description

Zimmer¿ Gender Solutions" Patellofemoral Joint Prosthesis Milling Handpiece

Reason for Recall

Complaints have been reported that the handpiece is inoperable. The investigation found that the complaints were related to the lack of preventative maintenance contrary to what is recommended in the instructions for use included with the handpiece. An inoperable device may lead to a delay in surgery

Distribution Pattern

Nationwide. CANADA HONG KONG KOREA SINGAPORE CHINA VENEZUELA AUSTRALIA JAPAN NETHERLANDS GERMANY INDIA TAIWAN

Code Information

Item Number 00-5927-040-00 Serial Number All