Z-0331-2024 Class II Ongoing
FDA device recall Z-0331-2024 was initiated by Aizu Olympus Co., Ltd. on Sep 25, 2023 and is designated Class II. Reason for recall: Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the… The recall status is ongoing. Affected quantity: 1243 units.
Recall details
- Product Type
- Devices
- Report Date
- Nov 29, 2023
- Initiation Date
- Sep 25, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1243 units
Product description
Cystonephroscope-endoscopic diagnosis and treatment within the bladder, urethra, and kidney Model Numbers: CYF-5, CYF-VH, CYF-VHR.
Reason for recall
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Distribution pattern
Worldwide - US Nationwide distribution.
Code information
Model Number (UDI-DI): CYF-5 (04953170292323), CYF-VH (04953170411250), CYF-VHR (04953170411298).
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0331-2024
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab
- Nov 11, 2026