Z-0418-2024 Class II Ongoing

Recalled by Aizu Olympus Co., Ltd. — Aizuwakamatsu

Recall Details

Product Type
Devices
Report Date
December 6, 2023
Initiation Date
October 16, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
24 incorrectly repaired units

Product Description

PCF-H190DL, EVIS EXERA III COLONOVIDEOSCOPE

Reason for Recall

Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.

Distribution Pattern

Domestic distribution to FL, KY, MA, NJ, NY, and PA

Code Information

UDI-DI 04953170363672 Serial Numbers 2942716, 2604497, 2502056