Z-0348-2024 Class II Ongoing
FDA device recall Z-0348-2024 was initiated by Aizu Olympus Co., Ltd. on Sep 25, 2023 and is designated Class II. Reason for recall: Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the… The recall status is ongoing.
Recall details
- Product Type
- Devices
- Report Date
- Nov 29, 2023
- Initiation Date
- Sep 25, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
Product description
Uretero-reno fiberscope-For endoscopic diagnosis and treatment within the ureter and kidney. Model Numbers: URF-P7, URF-P7R, URF-V2, URF-V2R, URF-V3, URF-V3R.
Reason for recall
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Distribution pattern
Worldwide - US Nationwide distribution.
Code information
Model Number (UDI-DI): URF-P7 (04953170403811), URF-P7R (04953170403835), URF-V2 (04953170343582), URF-V2R (04953170343612), URF-V3 (04953170403385), URF-V3R (04953170403392).
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0348-2024
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab