Z-0354-2014 Class II Terminated

Recalled by Restoration Robotics — San Jose, CA

FDA device recall Z-0354-2014 was initiated by Restoration Robotics on Oct 13, 2013 and is designated Class II. Reason for recall: Reports of discoloration (residue) on the pins of the skin tensioner have been received by several customers. Discoloration due to insufficient cleaning of stainless steel pins during passivation… The recall status is terminated (terminated Dec 10, 2013). Affected quantity: 515 US, 315 International (all packaging configurations).

Recall details

Product Type
Devices
Report Date
Nov 27, 2013
Initiation Date
Oct 13, 2013
Termination Date
Dec 10, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
515 US, 315 International (all packaging configurations)

Product description

ARTAS Disposable Kit ( Contains Skin Tensioner FP-45384; Inner Needle with Vacuum Tube FP-45385; Dissection Punch 4.0mm FP-23304; Saline Nose Cone FP-45387; Follicle Trap - FP-23306 and Non-sterile Disposable Clinical Kit FP -23267): and ARTAS Disposable Kits, 18 Gauge Epsilon (Contains Disposable Kit, 18 Gauge Epsilon FP 45547 and Non-Sterile Disposable Kit FP-23267). RX. Product Usage: Skin Tensioner is applied to patients scalp during follicular harvesting during hair restoration procedure. Product applies tension to scalp to facilitate follicular extractions.

Reason for recall

Reports of discoloration (residue) on the pins of the skin tensioner have been received by several customers. Discoloration due to insufficient cleaning of stainless steel pins during passivation process. The residue has been identified as a compound of iron and chlorine.

Distribution pattern

Worldwide Distribution - USA Nationwide in the countries of: Korea, Japan, Singapore, Thailand, United Kingdom, Canada, and Taiwan.

Code information

Model number FP-45616 and FP45173

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0354-2014
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Restoration Robotics? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Restoration Robotics: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Restoration Robotics →

Source: openFDA record for Z-0354-2014. Report a data issue: hello@regdatalab.com.