Z-0355-2014 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- November 27, 2013
- Initiation Date
- October 13, 2013
- Termination Date
- December 10, 2013
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 515 US, 315 International (all packaging cnfigurations)
Product Description
Sterile Disposable Kit ( Contains Skin Tensioner ASY-25067; Three Mark Needle Assembly ASY-25094; Nose Cone Assembly ASY-25098; Dissection 4 mm PRT-25117, follicle Trap PRT 23999): Product Usage: Skin Tensioner is applied to patients scalp during follicular harvesting during hair restoration procedure. Product applies tension to scalp to facilitate follicular extractions.
Reason for Recall
Reports of discoloration (residue) on the pins of the skin tensioner have been received by several customers. Discoloration due to insufficient cleaning of stainless steel pins during passivation process. The residue has been identified as a compound of iron and chlorine.
Distribution Pattern
Worldwide Distribution - USA Nationwide in the countries of: Korea, Japan, Singapore, Thailand, United Kingdom, Canada, and Taiwan.
Code Information
Model number FP-45547