Z-0364-2014 Class II Terminated

Recalled by Biomet, Inc. — Warsaw, IN

FDA device recall Z-0364-2014 was initiated by Biomet, Inc. on Oct 10, 2013 and is designated Class II. Reason for recall: Investigation identified that the design of the drill bit may make the suture anchor difficult to insert into the bone. The implant may skip across the hole or miss the hole during insertion. The recall status is terminated (terminated Jul 16, 2014). Affected quantity: 25.

Recall Details

Product Type
Devices
Report Date
Nov 27, 2013
Initiation Date
Oct 10, 2013
Termination Date
Jul 16, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
25

Product Description

Juggerknotless Drill Bit, REF 110003173, Sterile, Single Use Product Usage: The Juggerknotless Soft Anchors are indicated for soft tissue reattachment procedures in the shoulder, foot/ankle, hand/wrist, and hip.

Reason for Recall

Investigation identified that the design of the drill bit may make the suture anchor difficult to insert into the bone. The implant may skip across the hole or miss the hole during insertion.

Distribution Pattern

USA Nationwide Distribution in the states of VA, CA, Ky, and IN

Code Information

Catalog:110003173, Lot 127480, 185420, 233500, 233520