Z-0367-2025 Class II Ongoing

Recalled by Stryker Corporation — Portage, MI

Recall Details

Product Type
Devices
Report Date
November 20, 2024
Initiation Date
October 4, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,620 eaches

Product Description

Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Non-coated, 70mm - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-047-001

Reason for Recall

There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, Japan, Australia.

Code Information

GTIN: 07613327117219 Lot Number: 64117291