Z-0369-2025 Class II Ongoing

Recalled by Stryker Corporation — Portage, MI

FDA device recall Z-0369-2025 was initiated by Stryker Corporation on Oct 4, 2024 and is designated Class II. Reason for recall: There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in… The recall status is ongoing. Affected quantity: 21,490 eaches.

Recall Details

Product Type
Devices
Report Date
Nov 20, 2024
Initiation Date
Oct 4, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
21,490 eaches

Product Description

Stryker Neptune SafeAir Smoke Evacuation Pencil Non coated 70 mm Blade Push-Button - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-046-001

Reason for Recall

There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, Japan, Australia.

Code Information

GTIN: 07613327117196 Lot Numbers: 53867671 54409847 59053476 64855502 66494932

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0369-2025
openFDA file date
Oct 6, 2026
Loaded by RegDataLab
Oct 12, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it. FDA may have changed the recall since.