Z-0369-2025 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- November 20, 2024
- Initiation Date
- October 4, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 21,490 eaches
Product Description
Stryker Neptune SafeAir Smoke Evacuation Pencil Non coated 70 mm Blade Push-Button - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-046-001
Reason for Recall
There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Canada, Japan, Australia.
Code Information
GTIN: 07613327117196 Lot Numbers: 53867671 54409847 59053476 64855502 66494932