Z-0380-2017 Class II Terminated

Recalled by Stryker Sustainability Solutions — Tempe, AZ

FDA device recall Z-0380-2017 was initiated by Stryker Sustainability Solutions on Oct 6, 2016 and is designated Class II. Reason for recall: The EP Catheters may be mislabeled for French size during reprocessing. The recall status is terminated (terminated Dec 16, 2016). Affected quantity: 4 devices.

Recall Details

Product Type
Devices
Report Date
Nov 16, 2016
Initiation Date
Oct 6, 2016
Termination Date
Dec 16, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4 devices

Product Description

Reprocessed Daig Steerable and Daig Supreme Fixed Curve Diagnostic Electrophysiology Catheters Model Numbers: 401940, 402004 UPC Code: 00885825007409 00885825007744 Diagnostic electrophysiology (EP) catheters are specially designed electrode catheters that transmit electrical impulses and can be positioned for endocardial recording or stimulation. Diagnostic EP catheters incorporate a handpiece, a flexible shaft and a distal tip section containing diagnostic electrodes. Reprocessed Diagnostic EP Catheters are indicated for temporary intracardiac sensing, recording, stimulation, and electrophysiology mapping of cardiac structures.

Reason for Recall

The EP Catheters may be mislabeled for French size during reprocessing.

Distribution Pattern

U.S. distribution to the following; AZ, MN and MO.

Code Information

Item Number Serial Number Manufacturing Date 401940 1902898 3/31/2015 401940 1908621 3/31/2015 402004 1902392 1/9/2015 402004 1902389 1/9/2015

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0380-2017