Z-0394-2021 Class II Terminated

Recalled by Zimmer Biomet, Inc. — Warsaw, IN

FDA device recall Z-0394-2021 was initiated by Zimmer Biomet, Inc. on Sep 24, 2020 and is designated Class II. Reason for recall: Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse… The recall status is terminated (terminated May 20, 2022). Affected quantity: 412 units total.

Recall Details

Product Type
Devices
Report Date
Nov 11, 2020
Initiation Date
Sep 24, 2020
Termination Date
May 20, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
412 units total

Product Description

ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 55 mm Number: 00-8065-952-32

Reason for Recall

Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.

Distribution Pattern

US Nationwide

Code Information

All lots

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0394-2021
openFDA file date
Oct 6, 2026
Loaded by RegDataLab
Oct 12, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it. FDA may have changed the recall since.