Z-0399-2019 Class II Terminated

Recalled by Zimmer Biomet, Inc. — Warsaw, IN

FDA device recall Z-0399-2019 was initiated by Zimmer Biomet, Inc. on Sep 21, 2018 and is designated Class II. Reason for recall: There is a potential for comingling of the 43MM and 51 MM molds The recall status is terminated (terminated Apr 14, 2020). Affected quantity: 72 units.

Recall Details

Product Type
Devices
Report Date
Nov 14, 2018
Initiation Date
Sep 21, 2018
Termination Date
Apr 14, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
72 units

Product Description

StageOne Hip, Models 431207 & 431209Usage: Product Usage: The single-use cement spacer molds are sterile disposables made of medical grade silicone with a 316L stainless steel reinforcement stem. They are designed to be filled with polymethylmethacrylate/gentamicin bone cement, or equivalent, by injecting with a dispenser/gun into the mold. After the cement cures, the temporary spacers are to be removed from the molds with the reinforcement remaining as the core of the spacer, and placed into the joint space. The spacers remain in place (180 days or less) until the second stage of the two-stage procedure is performed to implant a conventional hip joint prosthesis.

Reason for Recall

There is a potential for comingling of the 43MM and 51 MM molds

Distribution Pattern

US Nationwide Distribution in the states of: AZ, FL, MI, OH, VA & WI

Code Information

Item Number 431207 Lot # 705550, UDI - (01)00880304447165(17)270927(10)705550 Item Number 431209 Lot # 705570, UDI - (01)00880304447141(17)270916(10)705570 Lot # 705580, UDI - (01)00880304447141(17)270916(10)705580

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0399-2019
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 24, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.