Z-0401-2020 Class II Terminated

Recalled by Orthofix Srl — Veronaverona, N/A

FDA device recall Z-0401-2020 was initiated by Orthofix Srl on Sep 27, 2018 and is designated Class II. Reason for recall: As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting… The recall status is terminated (terminated Feb 1, 2022).

Recall Details

Product Type
Devices
Report Date
Nov 20, 2019
Initiation Date
Sep 27, 2018
Termination Date
Feb 1, 2022
Voluntary/Mandated
Voluntary: Firm initiated

Product Description

ORTHOFIX Catalogue Number: ref: 99-93503US, ANKLE TRANSFIX PIN STERILE KIT, UDI: (01)18054242511304

Reason for Recall

As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.

Distribution Pattern

US: AZ, CA,FL,GA,HI,IA,IL,IN,KY,LA,MD,MI,MO,MS,NC,NY,OH,OK,OR,PA,SC,TN,TX,UT,VA,WA, WI, WV, Puerto Rico OUS: Kuwait, Australia, Saudi Arabia, United Arab Emirates, Ireland, Japan, Spain, Poland, Switzerland, Singapore, Peru, Brazil, Germany, Great Britain, France, Italy, Australia, South Africa, Belgium, Chile, The Netherlands, Lithuania, Sri Lanka, Denmark, Israel

Code Information

Lot #: B1200695, B1224714

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0401-2020
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 24, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.