Z-0431-2026 Class I Ongoing

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

Recall Details

Product Type
Devices
Report Date
November 19, 2025
Initiation Date
October 1, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4 system (1 US; 3 OUS)

Product Description

MAGNETOM Cima.X Upgrade. Model Number: 11689304.

Reason for Recall

There is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting paths are blocked, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

Distribution Pattern

US distribution to California. International distribution to Australia, Germany, Great Britian.

Code Information

Model Number: 11689304. UDI: (01)04056869975245(21)237304. Serial Number: 237304.