Z-0433-2013 Class II Terminated

Recalled by Sentec AG — Therwil, Baselland, N/A

FDA device recall Z-0433-2013 was initiated by Sentec AG on Nov 5, 2012 and is designated Class II. Reason for recall: A manufacturing defect was observed for the membrane changer insert that does not allow electrolyte to be present under the new membrane. The PCO2 performance of a sensor without electrolyte will be… The recall status is terminated (terminated Dec 18, 2012). Affected quantity: 195 boxes (9 per box).

Recall details

Product Type
Devices
Report Date
Nov 28, 2012
Initiation Date
Nov 5, 2012
Termination Date
Dec 18, 2012
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
195 boxes (9 per box)

Product description

SenTec V-Sign Membrane Changer (VS-MC), 9 pieces. The SenTec V-Sign Membrane Changer is indicated for use with the V-Sign Sensor only. The V-sign Sensor is indicated for use with the SenTec Digital Monitor when continuous non-invasive monitoring of carbon dioxide tension, oxygen saturation, and pulse rate are required for adult and pediatric patients.

Reason for recall

A manufacturing defect was observed for the membrane changer insert that does not allow electrolyte to be present under the new membrane. The PCO2 performance of a sensor without electrolyte will be deteriorated.

Distribution pattern

Distributed only in MO.

Code information

Lot numbers: 1CF5, 1D21, 1D60, 1DC6, 1DED, 1E0D

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0433-2013
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Sentec AG? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0433-2013. Report a data issue: hello@regdatalab.com.