Z-0433-2013 Class II Terminated

Recalled by Sentec AG — Therwil, Baselland, N/A

Recall Details

Product Type
Devices
Report Date
November 28, 2012
Initiation Date
November 5, 2012
Termination Date
December 18, 2012
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
195 boxes (9 per box)

Product Description

SenTec V-Sign Membrane Changer (VS-MC), 9 pieces. The SenTec V-Sign Membrane Changer is indicated for use with the V-Sign Sensor only. The V-sign Sensor is indicated for use with the SenTec Digital Monitor when continuous non-invasive monitoring of carbon dioxide tension, oxygen saturation, and pulse rate are required for adult and pediatric patients.

Reason for Recall

A manufacturing defect was observed for the membrane changer insert that does not allow electrolyte to be present under the new membrane. The PCO2 performance of a sensor without electrolyte will be deteriorated.

Distribution Pattern

Distributed only in MO.

Code Information

Lot numbers: 1CF5, 1D21, 1D60, 1DC6, 1DED, 1E0D