Z-1980-2023 Class II Ongoing
FDA device recall Z-1980-2023 was initiated by SenTec AG on May 2, 2023 and is designated Class II. Reason for recall: The sensors may experience an out-of-box failure because after recalibration, the sensors stayed in the software mode "production" and were not reset to "released" mode. The recall status is ongoing. Affected quantity: 11 in the U.S.
Recall details
- Product Type
- Devices
- Report Date
- Jun 28, 2023
- Initiation Date
- May 2, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 11 in the U.S.
Product description
Reason for recall
Distribution pattern
US Nationwide distribution in the states of CA, FL, IL, MA, MS, MT, NV, and TX.