Z-1980-2023 Class II Ongoing

Recalled by SenTec AG — Therwil

Recall Details

Product Type
Devices
Report Date
June 28, 2023
Initiation Date
May 2, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
11 in the U.S.

Product Description

V-Sign Sensor 2, REF VS-A/P/N, a component of Sentec Digital Monitoring System (model SDMS).

Reason for Recall

The sensors may experience an out-of-box failure because after recalibration, the sensors stayed in the software mode "production" and were not reset to "released" mode.

Distribution Pattern

US Nationwide distribution in the states of CA, FL, IL, MA, MS, MT, NV, and TX.

Code Information

Serial numbers 338947, 338952, 338953, 338963, 338967, 338989, 338991, 338993, 338994, 338996, 338997, UDI-DI 07640121880513.