Z-0450-2016 Class II Terminated
FDA device recall Z-0450-2016 was initiated by Biomet, Inc. on Nov 20, 2015 and is designated Class II. Reason for recall: These screws have a single lead (instead of a triple lead) thread. The screw may sit proud in the plate, If the screw does not lock into the plate; the screw may back out. If the screw sits proud… The recall status is terminated (terminated Sep 8, 2016). Affected quantity: 182, 965.
Recall Details
- Product Type
- Devices
- Report Date
- Dec 30, 2015
- Initiation Date
- Nov 20, 2015
- Termination Date
- Sep 8, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 182, 965
Product Description
A.L.P.S. 4.0mm Cancellous Locking Screws intended for use in stabilization and fixation of fractures, revision procedures, fusions, reconstructions (osteotomy) and non-unions of the bones.
Reason for Recall
These screws have a single lead (instead of a triple lead) thread. The screw may sit proud in the plate, If the screw does not lock into the plate; the screw may back out. If the screw sits proud, soft tissue irritation may occur. If the screw does not lock into the plate and backs out, a revision surgery may be necessary.
Distribution Pattern
AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR CHILE, COSTA RICA, NETHERLANDS, CANADA
Code Information
Part number and size (non-sterile) 816140010, 4.0X10MM; 816140012, 4.0X12MM; 816140014, 4.0X14MM; 816140016, 4.0X16MM; 816140018, 4.0X18MM; 816140020, 4.0X20MM; 816140022, 4.0X22MM; 816140024, 4.0X24MM; 816140026, 4.0X26MM; 816140028, 4.0X28MM; 816140030, 4.0X30MM; 816140032, 4.0X32MM; 816140034, 4.0X34MM; 816140036, 4.0X36MM; 816140038, 4.0X38MM; 816140040, 4.0X40MM; 816140042, 4.0X42MM; 816140044, 4.0X44MM; 816140046, 4.0X46MM; 816140048, 4.0X48MM; 816140050, 4.0X50MM; 816140055, 4.0X55MM; 816140060, 4.0X60MM; 816140065, 4.0X65MM; 816140070, 4.0X70MM; 816140075, 4.0X75MM; 816140080, 4.0X80MM; 816140085, 4.0X85MM; 816140090, 4.0X90MM; 816140095, 4.0X95MM; 816140100, 4.0X100MM Part number and size (sterile) 856140010, 4.0X10MM; 856140012, 4.0X12MM; 856140014, 4.0X14MM; 856140016, 4.0X16MM; 856140018, 4.0X18MM; 856140020, 4.0X20MM; 856140022, 4.0X22MM; 856140024, 4.0X24MM; 856140026, 4.0X26MM; 856140028, 4.0X28MM; 856140030, 4.0X30MM; 856140032, 4.0X32MM; 856140034, 4.0X34MM; 856140036, 4.0X36MM; 856140038, 4.0X38MM; 856140040, 4.0X40MM; 856140042, 4.0X42MM; 856140044, 4.0X44MM; 856140046, 4.0X46MM; 856140048, 4.0X48MM; 856140050, 4.0X50MM; 856140055, 4.0X55MM; 856140060, 4.0X60MM; 856140065, 4.0X65MM; 856140070, 4.0X70MM. All lots manufactured between 06/01/08 and 06/01/15.