Z-0450-2016 Class II Terminated

Recalled by Biomet, Inc. — Warsaw, IN

FDA device recall Z-0450-2016 was initiated by Biomet, Inc. on Nov 20, 2015 and is designated Class II. Reason for recall: These screws have a single lead (instead of a triple lead) thread. The screw may sit proud in the plate, If the screw does not lock into the plate; the screw may back out. If the screw sits proud… The recall status is terminated (terminated Sep 8, 2016). Affected quantity: 182, 965.

Recall Details

Product Type
Devices
Report Date
Dec 30, 2015
Initiation Date
Nov 20, 2015
Termination Date
Sep 8, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
182, 965

Product Description

A.L.P.S. 4.0mm Cancellous Locking Screws intended for use in stabilization and fixation of fractures, revision procedures, fusions, reconstructions (osteotomy) and non-unions of the bones.

Reason for Recall

These screws have a single lead (instead of a triple lead) thread. The screw may sit proud in the plate, If the screw does not lock into the plate; the screw may back out. If the screw sits proud, soft tissue irritation may occur. If the screw does not lock into the plate and backs out, a revision surgery may be necessary.

Distribution Pattern

AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR CHILE, COSTA RICA, NETHERLANDS, CANADA

Code Information

Part number and size (non-sterile) 816140010, 4.0X10MM; 816140012, 4.0X12MM; 816140014, 4.0X14MM; 816140016, 4.0X16MM; 816140018, 4.0X18MM; 816140020, 4.0X20MM; 816140022, 4.0X22MM; 816140024, 4.0X24MM; 816140026, 4.0X26MM; 816140028, 4.0X28MM; 816140030, 4.0X30MM; 816140032, 4.0X32MM; 816140034, 4.0X34MM; 816140036, 4.0X36MM; 816140038, 4.0X38MM; 816140040, 4.0X40MM; 816140042, 4.0X42MM; 816140044, 4.0X44MM; 816140046, 4.0X46MM; 816140048, 4.0X48MM; 816140050, 4.0X50MM; 816140055, 4.0X55MM; 816140060, 4.0X60MM; 816140065, 4.0X65MM; 816140070, 4.0X70MM; 816140075, 4.0X75MM; 816140080, 4.0X80MM; 816140085, 4.0X85MM; 816140090, 4.0X90MM; 816140095, 4.0X95MM; 816140100, 4.0X100MM Part number and size (sterile) 856140010, 4.0X10MM; 856140012, 4.0X12MM; 856140014, 4.0X14MM; 856140016, 4.0X16MM; 856140018, 4.0X18MM; 856140020, 4.0X20MM; 856140022, 4.0X22MM; 856140024, 4.0X24MM; 856140026, 4.0X26MM; 856140028, 4.0X28MM; 856140030, 4.0X30MM; 856140032, 4.0X32MM; 856140034, 4.0X34MM; 856140036, 4.0X36MM; 856140038, 4.0X38MM; 856140040, 4.0X40MM; 856140042, 4.0X42MM; 856140044, 4.0X44MM; 856140046, 4.0X46MM; 856140048, 4.0X48MM; 856140050, 4.0X50MM; 856140055, 4.0X55MM; 856140060, 4.0X60MM; 856140065, 4.0X65MM; 856140070, 4.0X70MM. All lots manufactured between 06/01/08 and 06/01/15.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0450-2016
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.