Z-0452-2022 Class II Terminated

Recalled by Biomeme, Inc. — Philadelphia, PA

FDA device recall Z-0452-2022 was initiated by Biomeme, Inc. on Oct 20, 2021 and is designated Class II. Reason for recall: The current IFU specifies a shelf-life of one week after resuspension of the RPC Buffer when stored at room temperature. Ongoing stability studies have indicated the need to shorten the life of the… The recall status is terminated (terminated Mar 5, 2024). Affected quantity: 11,607.

Recall details

Product Type
Devices
Report Date
Jan 12, 2022
Initiation Date
Oct 20, 2021
Termination Date
Mar 5, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
11,607

Product description

Biomeme SARS-CoV-2- Real Time RT-PCR Test, Instructions for Use, Part number 100002824 Subcomponent of: SARS-CoV-2 Go-Strips, 3000555; SARS-CoV-2 Go-Plates, 3000562; SARS-Cov-2 Bulk Vials, 3000564

Reason for recall

The current IFU specifies a shelf-life of one week after resuspension of the RPC Buffer when stored at room temperature. Ongoing stability studies have indicated the need to shorten the life of the RPC Buffer once resuspended from the current instruction of one week to two days when stored at room temperature (15degC-30degC).

Distribution pattern

Domestic distribution to AZ, CA, FL, HI, IN, KY, MD, MO, NC, NJ, NY, OH, OK, PA, TX, VA. Foreign distribution to Ghana, Dominican Republic, and Singapore.

Code information

All lots. Last lot of Kit distributed with the old IFU is 20210914.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0452-2022
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0452-2022. Report a data issue: hello@regdatalab.com.