Z-0513-2021 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- December 9, 2020
- Initiation Date
- October 23, 2020
- Termination Date
- January 25, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 105
Product Description
The Biomeme SARS-CoV-2 Real-Time RT-PCR Test is a molecular in vitro diagnostic test that is based on widely used nucleic acid amplification technology. The Biomeme SARS-CoV-2 Real-Time RT-PCR Test contains primers and probes and internal controls used in RT-PCR for the in vitro qualitative detection of SARS-CoV-2 RNA in upper respiratory specimens. The Biomeme SARS-CoV-2 Real-Time RT-PCR Test utilizes Biomemes M1 Sample Prep Cartridge for RNA extraction, Biomemes SARS-CoV-2 Go-Strips assay, and Biomemes portable Franklin" Real-Time qPCR Thermocycler. Franklins companion mobile app, Biomeme Go, scans tests, runs PCR experiments online or offline, and is used to quickly interpret your test results while conveniently syncing data to the Biomeme Cloud. Biomeme: 1000013 (Android Smartphone w/ Biomeme Go Mobile App) and Biomeme: 1000012 (Rugged Android Smartphone w/ Biomeme Go Mobile App) are controller for Biomeme Franklin thermocycler.
Reason for Recall
The firm has become aware of nine reports by a single customer that the software made false positive calls. The investigation determined that the root cause is the higher than expected amber-to-red co-excitation in the assay.
Distribution Pattern
Worldwide distribution - US Nationwide distribution including in the states of CA, IN, NY, OH, PA, and the countries of Canada, Singapore, Zimbabwe, and Dominican Republic.
Code Information
Revision 2020-03-24