Z-0457-2019 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-0457-2019 was initiated by Siemens Medical Solutions USA, Inc on Oct 19, 2018 and is designated Class II. Reason for recall: Vibrations of the C-arm of the plane B may result in reduced image quality caused by the screw connection to the C-arm in the ceiling support The recall status is terminated (terminated Apr 27, 2020). Affected quantity: 14 units.

Recall Details

Product Type
Devices
Report Date
Nov 21, 2018
Initiation Date
Oct 19, 2018
Termination Date
Apr 27, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
14 units

Product Description

ARTIS zee biplane systems with Material 10094141- a Angiographic x-ray system Product Usage: Artis zee is a dedicated angiography system developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee includes cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee can also support the acquisition of position triggered imaging for spatial data synthesis.

Reason for Recall

Vibrations of the C-arm of the plane B may result in reduced image quality caused by the screw connection to the C-arm in the ceiling support

Distribution Pattern

US Nationwide Distribution

Code Information

Serial Numbers: Serial 155122 155149 155151 155105 155148 154671 155126 155131 155132 155119 155121 155118 155147

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0457-2019
openFDA file date
Oct 21, 2026
Loaded by RegDataLab
Oct 23, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.