Z-0461-2016 Class II Terminated

Recalled by Siemens Healthcare Diagnostics Inc. — Brookfield, CT

FDA device recall Z-0461-2016 was initiated by Siemens Healthcare Diagnostics Inc. on Nov 10, 2015 and is designated Class II. Reason for recall: Siemens has identified RMS refrigeration compressors that are missing the safety cover on the electrical termination block. The exposed electrical hazard on the electrical termination block is… The recall status is terminated (terminated Aug 23, 2017). Affected quantity: 151.

Recall Details

Product Type
Devices
Report Date
Dec 30, 2015
Initiation Date
Nov 10, 2015
Termination Date
Aug 23, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
151

Product Description

Reagent Management System (RMS) Dimension RxL Max HM STM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension¿ clinical chemistry system. The RMS doubles the available storage of refrigerated Flex¿ reagent cartridges on the Dimension¿ system from 44 to 88. The RMS loads, hydrates, and automatically removes Flex¿ reagent cartridges, allowing the Dimension¿ system to process samples without interruption.

Reason for Recall

Siemens has identified RMS refrigeration compressors that are missing the safety cover on the electrical termination block. The exposed electrical hazard on the electrical termination block is located directly behind the RMS Waste Container.

Distribution Pattern

US Nationwide Distribution: FL, NY, SC, TX, LA, AZ, WI, IA, IL, OH, AL, AK, NH, OK, MD, CA, NM, MS, KY, NJ, MN, GA, CO, MA, SD, TN, MO, PA, MT, SC, NE, KS, VA, AL, NV, WY, OR, PR, MO, OK, GU, HI, AR, ME, VT, CT, NC and the countries of Canada and Mexico.

Code Information

Dimension RXL MAX HM STM Catalog Numbers 752000.941 (Domestic)/752000.951 (Intl). RMS for RxL Catalog Numbers 753000.901 (Domestic)/753000.902 (Intl). All serial numbers.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0461-2016
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.