Z-0463-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- December 14, 2022
- Initiation Date
- November 22, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- N/A
Product Description
Philips Upgrade to MR 7700 System, Model Number 782130
Reason for Recall
Gradient Coil may act as a heat source with a potential to produce smoke and/or fire.
Distribution Pattern
US Nationwide.
Code Information
Model 782120: UDI-DI: 00884838104402; Serial Numbers: 45298 45301