Z-0472-2016 Class II Terminated

Recalled by Biomet, Inc. — Warsaw, IN

FDA device recall Z-0472-2016 was initiated by Biomet, Inc. on Dec 2, 2015 and is designated Class II. Reason for recall: There may be a separation and an insufficient weld between the spring and washer. The weld between the spring and washer may fracture, causing the Mamba Suture Passer instruments jaw to not actuate… The recall status is terminated (terminated Oct 27, 2016). Affected quantity: 13.

Recall Details

Product Type
Devices
Report Date
Dec 30, 2015
Initiation Date
Dec 2, 2015
Termination Date
Oct 27, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13

Product Description

Green Mamba Suture Passer

Reason for Recall

There may be a separation and an insufficient weld between the spring and washer. The weld between the spring and washer may fracture, causing the Mamba Suture Passer instruments jaw to not actuate properly. 6 complaints have been received in which 3 have been reported with a delay up to 10 minutes.

Distribution Pattern

FL, NJ, WI, UT, TX, KY, KS, OH, GA, LA, AL, NY, NC, AR, MO, VA, CA, MN, IN, HI

Code Information

Catalog Number: 110010850 Lot Number Identification: 167890, 231120, 253210, 550050, 551580, 716300, 953230

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0472-2016
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.