Z-0472-2023 Class II Ongoing

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-0472-2023 was initiated by Siemens Medical Solutions USA, Inc on Nov 9, 2022 and is designated Class II. Reason for recall: When selecting the site fraction group for treatment, the User may select the wrong site for treatment, and potentially deliver the dose to wrong isocenter The recall status is ongoing. Affected quantity: 8 units US.

Recall Details

Product Type
Devices
Report Date
Dec 21, 2022
Initiation Date
Nov 9, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8 units US

Product Description

ARTISTE, ONCOR and PRIMUS with syngo RT Therapist with software version 4.3.1 MR3. Material Number: 8162815 Intended Use: Family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer.

Reason for Recall

When selecting the site fraction group for treatment, the User may select the wrong site for treatment, and potentially deliver the dose to wrong isocenter

Distribution Pattern

US Nationwide distribution in the states of CA, IA, LA, NE, NV, OH, SC.

Code Information

UDI: N/A S/N: 10736, 10737, 10822, 10850, 10886,10843, 10865,10025 Software version 4.3.1 MR3

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0472-2023
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 24, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.