Z-0474-2013 Class II Terminated

Recalled by Biomet, Inc. — Warsaw, IN

FDA device recall Z-0474-2013 was initiated by Biomet, Inc. on Nov 1, 2012 and is designated Class II. Reason for recall: The items in this lot are missing the ion implantation feature. The surface of the head may be softer and more prone to scratching, which in tum could potentially cause higher wear of the poly… The recall status is terminated (terminated Aug 29, 2013). Affected quantity: 4.

Recall Details

Product Type
Devices
Report Date
Dec 12, 2012
Initiation Date
Nov 1, 2012
Termination Date
Aug 29, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4

Product Description

Bio-Modular Humeral Head 48mm x 24mm Ti 6AI4V/Ion Implanted (Orthopedic Implant) Product Usage: The OSS Knee Reamer Sleeves are primarily used to guide the bullet tip of cylindrical reamers during the tibial preparation in OSS procedures. They are reusable.

Reason for Recall

The items in this lot are missing the ion implantation feature. The surface of the head may be softer and more prone to scratching, which in tum could potentially cause higher wear of the poly bearing. Increased wear could potentially lead to a shorter implant life.

Distribution Pattern

USA Nationwide Distribution in the state of WI.

Code Information

Catalog Number: 113740 and Lot Number Identification: 779270

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0474-2013
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.