Z-0477-2019 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- November 28, 2018
- Initiation Date
- July 25, 2018
- Termination Date
- April 27, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 108
Product Description
Binding Site Human IgM Kit (for SPAPLUS), REF: NK012.S The device consists of reaction buffer, sheep anti-IgM antiserum, six calibrators and two controls. The calibrators and controls are based on stabilized human serum. All components contain sodium azide as a preservative and are provided in ready-to-use liquid format.
Reason for Recall
After running the calibration curve, the recovery of the IgM kit control results decrease demonstrating a reduction in accuracy within the same day; control results may be out of range one day after running the assay calibration curve. Samples have been shown to be similarly affected.
Distribution Pattern
U.S.: IL, WV, NE, SD, NY, IN, WA, MD, NJ, PA, NC, AR, FL, VA
Code Information
Lot: 407483-1 and 407483-2