Z-0477-2019 Class II Terminated

Recalled by The Binding Site Group, Ltd. — Birmingham, N/A

Recall Details

Product Type
Devices
Report Date
November 28, 2018
Initiation Date
July 25, 2018
Termination Date
April 27, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
108

Product Description

Binding Site Human IgM Kit (for SPAPLUS), REF: NK012.S The device consists of reaction buffer, sheep anti-IgM antiserum, six calibrators and two controls. The calibrators and controls are based on stabilized human serum. All components contain sodium azide as a preservative and are provided in ready-to-use liquid format.

Reason for Recall

After running the calibration curve, the recovery of the IgM kit control results decrease demonstrating a reduction in accuracy within the same day; control results may be out of range one day after running the assay calibration curve. Samples have been shown to be similarly affected.

Distribution Pattern

U.S.: IL, WV, NE, SD, NY, IN, WA, MD, NJ, PA, NC, AR, FL, VA

Code Information

Lot: 407483-1 and 407483-2