Z-0477-2019 Class II Terminated

Recalled by The Binding Site Group, Ltd. — Birmingham, N/A

FDA device recall Z-0477-2019 was initiated by The Binding Site Group, Ltd. on Jul 25, 2018 and is designated Class II. Reason for recall: After running the calibration curve, the recovery of the IgM kit control results decrease demonstrating a reduction in accuracy within the same day; control results may be out of range one day after… The recall status is terminated (terminated Apr 27, 2021). Affected quantity: 108.

Recall details

Product Type
Devices
Report Date
Nov 28, 2018
Initiation Date
Jul 25, 2018
Termination Date
Apr 27, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
108

Product description

Binding Site Human IgM Kit (for SPAPLUS), REF: NK012.S The device consists of reaction buffer, sheep anti-IgM antiserum, six calibrators and two controls. The calibrators and controls are based on stabilized human serum. All components contain sodium azide as a preservative and are provided in ready-to-use liquid format.

Reason for recall

After running the calibration curve, the recovery of the IgM kit control results decrease demonstrating a reduction in accuracy within the same day; control results may be out of range one day after running the assay calibration curve. Samples have been shown to be similarly affected.

Distribution pattern

U.S.: IL, WV, NE, SD, NY, IN, WA, MD, NJ, PA, NC, AR, FL, VA

Code information

Lot: 407483-1 and 407483-2

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0477-2019
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for The Binding Site Group, Ltd.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0477-2019. Report a data issue: hello@regdatalab.com.