Z-0489-2025 Class II Ongoing

Recalled by Roche Diagnostics Operations, Inc. — Indianapolis, IN

Recall Details

Product Type
Devices
Report Date
November 27, 2024
Initiation Date
October 17, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
458 units

Product Description

cobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, in vitro test for total L-homocysteine

Reason for Recall

Abnormal calibration signals and control recovery issues on the Homocysteine Enzymatic (HCYS) Assay can produce erroneous patient results (positive or negative bias)

Distribution Pattern

US: NJ, TX, NY, FL, IN, TX, OK, SC, IA,

Code Information

UDI/DI 04015630929979, Lot Number 747029, exp. 30-SEPT-2024