Z-1009-2026 Class II Ongoing
FDA device recall Z-1009-2026 was initiated by Roche Diagnostics Operations, Inc. on Nov 21, 2025 and is designated Class II. Reason for recall: Issues identified: 1) Results vary by Instrument Platform; 2) Calibrator Lot-to-Lot Variability. Issues may lead to delayed or incorrect diagnosis, as well as delay of treatment or incorrect… The recall status is ongoing. Affected quantity: 1300 cassettes.
Recall details
- Product Type
- Devices
- Report Date
- Jan 14, 2026
- Initiation Date
- Nov 21, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1300 cassettes
Product description
Reason for recall
Distribution pattern
US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, ME, MI, MN, NC, NJ, NM, NY, OH, PA, PR, TN, TX, UT, WA, WV.