Z-0491-2019 Class II Terminated

Recalled by Becton Dickinson & Company — Franklin Lakes, NJ

FDA device recall Z-0491-2019 was initiated by Becton Dickinson & Company on Aug 30, 2018 and is designated Class II. Reason for recall: BD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when used with the Sysmex UF Series Analyzers. BD recommends following Good Clinical Practice… The recall status is terminated (terminated Oct 21, 2020). Affected quantity: Total: 29,982,150 BD cups (US); 97,800,900 BD cups (OUS).

Recall Details

Product Type
Devices
Report Date
Nov 28, 2018
Initiation Date
Aug 30, 2018
Termination Date
Oct 21, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
Total: 29,982,150 BD cups (US); 97,800,900 BD cups (OUS)

Product Description

BD Vacutainer Urine Collection Cups, Cat. No. 364941

Reason for Recall

BD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when used with the Sysmex UF Series Analyzers. BD recommends following Good Clinical Practice (GCP) by using microscopy to confirm an elevated RBC Count in a Urine sample as per CLSI GP16-3A. 29(4) and CAP Checklist on Urinalysis. Product will continue to ship to not impact critical patient care.

Distribution Pattern

Illinois

Code Information

All lots within expiry

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0491-2019
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 24, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.