Z-0491-2025 Class II Ongoing

Recalled by Abbott Medical — Saint Paul, MN

FDA device recall Z-0491-2025 was initiated by Abbott Medical on Oct 17, 2024 and is designated Class II. Reason for recall: A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the… The recall status is ongoing. Affected quantity: 1 US, 9 OUS.

Recall details

Product Type
Devices
Report Date
Nov 27, 2024
Initiation Date
Oct 17, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 US, 9 OUS

Product description

Abbott Navitor Titan, Vision Technology, Transcatheter Aortic Heart Valve, 35 mm, NVRO-35, Sterile

Reason for recall

A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or cause potential valve failure.

Distribution pattern

New York, Australia, Belgium, Germany, Iran, Israel, Japan, Kazakhstan, Oman, Poland, Romania, Slovakia, Spain, Turkey, United Kingdom.

Code information

UDI-DI (GTIN): 05415067045768, Serial Numbers: 20078589, 20357219, 20357226, 20357685, 20358319, 20358343, 20358351, 20358363, 20358368, & 20358370.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0491-2025
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Abbott Medical? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0491-2025. Report a data issue: hello@regdatalab.com.