Z-0494-2025 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- November 27, 2024
- Initiation Date
- October 17, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 8 OUS
Product Description
Abbott Navitor, Vision Technology, Transcatheter Aortic Heart Valve, 27 mm, NVRO-27, Sterile
Reason for Recall
A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or cause potential valve failure.
Distribution Pattern
New York, Australia, Belgium, Germany, Iran, Israel, Japan, Kazakhstan, Oman, Poland, Romania, Slovakia, Spain, Turkey, United Kingdom.
Code Information
UDI-DI (GTIN): 5415067045799, Serial Numbers: 20158135, 20172203, 20183968, 20184552, 20184670, 20184675, 20190249, & 20213055.