Z-0532-2021 Class II Terminated

Recalled by Canon Medical System, USA, INC. — Tustin, CA

FDA device recall Z-0532-2021 was initiated by Canon Medical System, USA, INC. on Oct 8, 2020 and is designated Class II. Reason for recall: There is a potential that the cable of the Flat panel detector may break during a procedure which may cause the x-ray device to stop before the examination is completed. This could result in a delay… The recall status is terminated (terminated Dec 6, 2022). Affected quantity: 14 units.

Recall details

Product Type
Devices
Report Date
Dec 9, 2020
Initiation Date
Oct 8, 2020
Termination Date
Dec 6, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
14 units

Product description

CAS-930A Flat Panel Monitor (FPD) used in conjunction with the Canon Interventional Angiography System, Model numbers: Infinix-8000C and Infinix-8000H - Product Usage: This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

Reason for recall

There is a potential that the cable of the Flat panel detector may break during a procedure which may cause the x-ray device to stop before the examination is completed. This could result in a delay with procedure and/or potential damage to the patients blood vessels due to removal without visibility.

Distribution pattern

US Nationwide distribution including in the states of LA, CA, CO, WV, GA, NH, FL, PA, WI, MN, ID.

Code information

Affected Serial Numbers: 1 AFA1662001, 2 AFB1872005, 3 AFB1832003, 4 AFB1872004, 5 AFB1892006, 6 BFA18Y7001, 7 AEA17X2001, 8 AFA1692002, 9 AEB1812002, 10 BFA1932001, 11 BFB1952003, 12 BFB1942002, 13 BFB1972004, 14 BFB1972005, 15 AFB1892007, 16 BEC2032001, 17 BFC2062007, 18 BFC2062008, 19 BHA1992001, 20 BFC2012006

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0532-2021
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Canon Medical System, USA, INC.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Canon Medical System, USA, INC.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Canon Medical System, USA, INC. →

Source: openFDA record for Z-0532-2021. Report a data issue: hello@regdatalab.com.