Z-0533-2013 Class II Terminated
FDA device recall Z-0533-2013 was initiated by Biomet, Inc. on Nov 15, 2012 and is designated Class II. Reason for recall: Biomet investigation identified that the square end of the offset tibial tray adaptor locking insert is oversized and may not engage insert tool screwdriver. This could cause a possible delay in… The recall status is terminated (terminated Nov 8, 2013). Affected quantity: 200.
Recall Details
- Product Type
- Devices
- Report Date
- Dec 26, 2012
- Initiation Date
- Nov 15, 2012
- Termination Date
- Nov 8, 2013
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 200
Product Description
Biomet Offset Tibial Tray 2.5 mm Adaptor, REF 141490 Sterile, QTY.1 Product Usage: Knee joint replacement prostheses intended for application with or without bone cement. Knee joint replacement components include femoral, tibial, and patellar components, intended for both primary and revision applications. Indications: 1) painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis when one or more compartments are involved, 2) correction of varus, valgus, or posttraumatic deformity, 3) correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure
Reason for Recall
Biomet investigation identified that the square end of the offset tibial tray adaptor locking insert is oversized and may not engage insert tool screwdriver. This could cause a possible delay in surgery greater than 30 minutes, potentially exposing the patient to the increased risks from being under anesthesia for a longer duration.
Distribution Pattern
Worldwide distribution - USA (nationwide) and the countries of Australia, Canada, Chile, Europe, Japan, Korea, Panama, Puerto Rico and Thailand.
Code Information
Catalog Number: 141490 Lot 378330,378320,407230,485010,505140,485000,506070,533400,54048, 439700,797820