Z-0533-2013 Class II Terminated

Recalled by Biomet, Inc. — Warsaw, IN

FDA device recall Z-0533-2013 was initiated by Biomet, Inc. on Nov 15, 2012 and is designated Class II. Reason for recall: Biomet investigation identified that the square end of the offset tibial tray adaptor locking insert is oversized and may not engage insert tool screwdriver. This could cause a possible delay in… The recall status is terminated (terminated Nov 8, 2013). Affected quantity: 200.

Recall Details

Product Type
Devices
Report Date
Dec 26, 2012
Initiation Date
Nov 15, 2012
Termination Date
Nov 8, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
200

Product Description

Biomet Offset Tibial Tray 2.5 mm Adaptor, REF 141490 Sterile, QTY.1 Product Usage: Knee joint replacement prostheses intended for application with or without bone cement. Knee joint replacement components include femoral, tibial, and patellar components, intended for both primary and revision applications. Indications: 1) painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis when one or more compartments are involved, 2) correction of varus, valgus, or posttraumatic deformity, 3) correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure

Reason for Recall

Biomet investigation identified that the square end of the offset tibial tray adaptor locking insert is oversized and may not engage insert tool screwdriver. This could cause a possible delay in surgery greater than 30 minutes, potentially exposing the patient to the increased risks from being under anesthesia for a longer duration.

Distribution Pattern

Worldwide distribution - USA (nationwide) and the countries of Australia, Canada, Chile, Europe, Japan, Korea, Panama, Puerto Rico and Thailand.

Code Information

Catalog Number: 141490 Lot 378330,378320,407230,485010,505140,485000,506070,533400,54048, 439700,797820

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0533-2013
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.