Z-0539-2019 Class II Terminated
FDA device recall Z-0539-2019 was initiated by Zimmer Biomet, Inc. on Oct 22, 2018 and is designated Class II. Reason for recall: Various Trauma Guide Wires- creases in the sealing area on the end opposite of the product's chevron opening compromising product sterility The recall status is terminated (terminated Dec 30, 2020). Affected quantity: 3764.
Recall Details
- Product Type
- Devices
- Report Date
- Dec 12, 2018
- Initiation Date
- Oct 22, 2018
- Termination Date
- Dec 30, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3764
Product Description
Zimmer Natural Nail¿ System - Tear Drop Guide Wire, 2.4 mm Diameter, 100 cm Length Item Number: 47-2490-097-01 Product Usage: A single use surgical instrument used during orthopaedic surgery for implantation of a prosthesis.
Reason for Recall
Various Trauma Guide Wires- creases in the sealing area on the end opposite of the product's chevron opening compromising product sterility
Distribution Pattern
Worldwide - US Nationwide Distribution Foreign: CANADA AUSTRALIA BRAZIL CHILE HONG KONG JAPAN MALAYSIA NETHERLANDS NICARAGUA SINGAPORE TAIWAN
Code Information
All lots expiring prior to September 30, 2023