Z-0551-2013 Class II Terminated

Recalled by Zimmer, Inc. — Warsaw, IN

FDA device recall Z-0551-2013 was initiated by Zimmer, Inc. on Nov 14, 2012 and is designated Class II. Reason for recall: All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and… The recall status is terminated (terminated Apr 16, 2014). Affected quantity: 189.

Recall Details

Product Type
Devices
Report Date
Dec 26, 2012
Initiation Date
Nov 14, 2012
Termination Date
Apr 16, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
189

Product Description

Persona, The Personalized Knee System, Tibial Sizing Plate, Cemented, Nonsterile, Size D Left; Size D Right This device is used with the Persona Knee System for patients with severe knee pain and disability due to: Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. o Collagen disorders, and/or avascular necrosis of the femoral condyle. o Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. o Moderate valgus, varus, or flexion deformities

Reason for Recall

All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and replaced with enhanced products and instruments to reduce risks of tibial perforation which could result in tibial bone fracture due to the stress riser created by a perforation; tibial plate loosening

Distribution Pattern

Worldwide Distribution - USA (nationwide) and internationally to Germany and Europe.

Code Information

Part Number 42-5399-067-01(left) Lot 62050138, 62121496 Part Number 42-5399-067-02 (right), Lot 62067442, 62075280, 62121527

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0551-2013
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.