Z-0569-2021 Class II Terminated

Recalled by Ra Medical Systems, Inc. — Carlsbad, CA

FDA device recall Z-0569-2021 was initiated by Ra Medical Systems, Inc. on October 16, 2020 and is designated Class II. Reason for recall: Due to footswitch not meeting required specification for protection from ingress of solids or liquids. If spilling of liquids near or on the footswitch, it may fail in the 'on' position and allow the… The recall status is terminated (terminated January 11, 2023). Affected quantity: 36 devices.

Recall Details

Product Type
Devices
Report Date
December 23, 2020
Initiation Date
October 16, 2020
Termination Date
January 11, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
36 devices

Product Description

DABRA Laser (RA-308 Excimer Laser) Intended Use: For use in ablating a channel in occlusive peripheral vascular disease. Indications: For crossing chronic total occlusions in patients with symptomatic infrainguinal lower extremity vascular disease.

Reason for Recall

Due to footswitch not meeting required specification for protection from ingress of solids or liquids. If spilling of liquids near or on the footswitch, it may fail in the 'on' position and allow the laser to continue to lase.

Distribution Pattern

US Distribution to states of: AL, AZ, CA, FL, IL, IN, KY, LA, MD, NC, NV, NY, OK, OR, TX, and WA; and to OUS country: Japan.

Code Information

Model Number RA-308 Product/Catalog Number: 1200-0001-01 Serial Numbers: RA00027, RA00046, RA00049, RA00050, RA00054, RA00056, RA00057, RA00058, RA00064, RA00069, RA00073, RA00081, RA00083, RA00084, RA00085, RA00089, RA00097, RA00098, RA00114, RA00115, RA00120, RA00121, RA00125, RA00127, RA00135, RA00138, RA00148, RA00153, RA00154, RA00160, RA00164, RA00165, RA00167, RA00170, RA00175, and RA00177.