Z-1631-2020 Class II Terminated

Recalled by Ra Medical Systems, Inc. — Carlsbad, CA

FDA device recall Z-1631-2020 was initiated by Ra Medical Systems, Inc. on Aug 23, 2019 and is designated Class II. Reason for recall: At doses of less than 100 mJ, in custom and auto repeat modes, a software error in the excimer laser phototherapy system causes the device to fail to cease firing as long as the foot pedal or… The recall status is terminated (terminated May 5, 2021). Affected quantity: 10.

Recall details

Product Type
Devices
Report Date
Apr 8, 2020
Initiation Date
Aug 23, 2019
Termination Date
May 5, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
10

Product description

Pharos Excimer Laser, Model: EX-308 - Product Usage: The Pharos EX-308 Excimer Laser System is designed to treat the skin disorders psoriasis, vitiligo, atopic dermatitis, and leukoderma using phototherapy. The excimer laser system produces collimated 308 nm radiation, within the invisible region of the electromagnetic spectrum. The UV radiation falls within the range of ultraviolet B (UVB), which is used to treat photo-responsive skin disorders.

Reason for recall

At doses of less than 100 mJ, in custom and auto repeat modes, a software error in the excimer laser phototherapy system causes the device to fail to cease firing as long as the foot pedal or handpiece switch is depressed. This malfunction can cause overtreatment, which could result in strong erythema, and possibly cause blisters.

Distribution pattern

U.S. Nationwide distribution in the states of CA, FL, OR, MS, NY.

Code information

All serial numbers.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1631-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Ra Medical Systems, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Ra Medical Systems, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Ra Medical Systems, Inc. →

Source: openFDA record for Z-1631-2020. Report a data issue: hello@regdatalab.com.