Z-1631-2020 Class II Terminated
FDA device recall Z-1631-2020 was initiated by Ra Medical Systems, Inc. on Aug 23, 2019 and is designated Class II. Reason for recall: At doses of less than 100 mJ, in custom and auto repeat modes, a software error in the excimer laser phototherapy system causes the device to fail to cease firing as long as the foot pedal or… The recall status is terminated (terminated May 5, 2021). Affected quantity: 10.
Recall details
- Product Type
- Devices
- Report Date
- Apr 8, 2020
- Initiation Date
- Aug 23, 2019
- Termination Date
- May 5, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 10
Product description
Reason for recall
Distribution pattern
U.S. Nationwide distribution in the states of CA, FL, OR, MS, NY.